3 MAINTENANCE
Fig. 19: Cell inlet filter, spare parts
12 Fit a new O-ring (black) on the outside.
Fig. 20: Cell inlet filter, O-ring, outer
13 Disconnect the nozzle on the cell with a 5 mm Allen key (see Fig. 17 “Cell inlet filter,
opened”, page 23).
14 Replace the spring washer of the nozzle.
15 Screw the nozzle tight.
16 Attach the cell inlet filter to the cell with new screws.
17 Fasten hose connections.
18 Close the cell cover and screw tight.
1Sintered metal washer
2Lock washer
3O-ring, white
4O-ring, black
5Cell inlet filter
Keep the nozzle for reassembly.
Observe the sequence:
First put the spring washer, then the washer on the screw.