8024638/AE00/V1-0/2019-09| SICKSERVICE MANUAL | MCS200HW
Subject to change without notice
3 MAINTENANCE
Fig. 41: Upper mirror plate
8 Remove the old O-ring from the mirror plate:
Carefully remove O-ring residues from the mirror plate and cell.
Protect the lower mirror plate against O-ring residues.
9 Fit a new O-ring on the mirror plate.
10 Fit the upper mirror plate.
1Mirror plate
2Coated surface
3O-ring
4Recesses for levering out
NOTE: Leak tightness of the system
Damaged surfaces impair the leak tightness of the system.
Do not damage surfaces.
Suitable aids for removing O-ring residues:
Flat tool (slot screwdriver, scalpel or similar)
–Acetone
NOTE: Material damage
Protect mirror plates from insulation particles. Do not touch the insulation material.
Observe alignment: Align the labeled side of the mirror plate to the front of the cell inlet
filter.