8024638/AE00/V1-0/2019-09| SICK S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
MAINTENANCE 3
4 Remove the old O-rings (3 pieces each):
Carefully remove O-ring residues from the tubes and cell.
1 Remove O-rings 2 and 3 from the cell.
Fig. 50: Tube, O-rings 2 and 3
2 Remove O-ring 1 from the inside of the tube:
a) Remove the crystal glass from the tube:
Use a tool (e.g. a hexagon key) to press the crystal glass out of the tube from below.
>>>
This could destroy the crystal glass.
b) Remove O-ring 1.
Fig. 51: Removing the tube crystal glass
5 Fit new O-rings and new crystal glass (observe sequence):
1 Insert O-ring 1 inside the tube.
2 Place 0-ring 2 outside around the tube.
NOTE: Leak tightness of the system
Damaged surfaces impair the leak tightness of the system.
Do not damage surfaces.
Suitable aids for removing O-ring residues:
Flat tool (slot screwdriver, scalpel or similar)
–Acetone
1O-ring 2
2O-ring 3
O-Ring 1 is located under the crystal glass of the tube and probably sticks to the crystal
glass when removed.

Table of Contents

Other manuals for SICK MCS200HW

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SICK MCS200HW and is the answer not in the manual?

SICK MCS200HW Specifications

General IconGeneral
BrandSICK
ModelMCS200HW
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals