5 REPAIRS
9 Close the cell cover and screw tight.
10 Adjust the oxygen sensor.
1 Select Parameterization -> O2-sensor.
2 Check nominal value: The nominal value must be 20.95%.
3 Select Adjustment -> Start adjustment.
4 Click Adjust O2.
>>>
Oxygen sensor is adjusted.
In the case of the maintenance message S047 Deviation O2 Just. too large due to
exceeding the limit value, the message can be reset under Maintenance ->
Acknowledge active message.