8024638/AE00/V1-0/2019-09| SICK S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
REPAIRS 5
5 Loosen the fastening screws on the lower mirror plate with a 5 mm Allen key (6 pieces).
.
Fig. 106: Fastening screws, lower mirror plate
6 Lever out the mirror plate:
1 Screw the screws, without spring washers, into the thread openings provided on the
mirror plate (2 pieces).
Fig. 107: Thread openings, lower mirror plate
2 Tighten the screws with a 5 mm Allen key.
▸ Tighten fastening screws parallel.
▸ Ensure the screws grip properly and avoid thread seizure.
7 Remove the lower mirror plate.
1Fastening screws
1Thread openings
Alternative:
– Unscrew and remove all cell covers (see Chapter “Removing the complete cell”,
page 65).
– Apply the flat side of a screwdriver to the side recesses of the mirror plate.
– Lever out the mirror plate.