Environmental Specifications
CE Electrical Equipment Requirements
For information regarding the equipment needed to provide the electrical
service listed under “Service Requirements” at the end of Chapter 3, please
refer to specification EN-309, “Plug, Outlet and Socket Couplers for Indus-
trial Uses,” listed in the official Journal of the European Communities.
Environmental Specifications
The environmental conditions under which the laser system will function
are listed below:
Indoor use
Altitude: up to 2000 m
Temperatures: 10° C to 40° C
Maximum relative humidity: 80% non-condensing for temperatures up to
31° C.
Mains supply voltage: do not exceed ±10% of the nominal voltage
Insulation category: II
Pollution degree: 2
FCC Regulations
This equipment has been tested and found to comply with the limits for a
Class A digital device pursuant to Part 15 of the FCC Rules. These limits
are designed to provide reasonable protection against harmful interference
when the equipment is operated in a commercial environment. This equip-
ment generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instruction manual, may cause
harmful interference to radio communications. Operation of this equipment
in a residential area is likely to cause harmful interference in which case
the user will be required to correct the interference at his own expense.
Modifications to the laser system not expressly approved by Spectra-
Physics could void your right to operate the equipment.