Tsunami Mode-Locked Ti:sapphire Laser
4-2
Input Bezel Connections
The input bezel connections attach to the Model 3910 purge unit (optional
on the Model 3941), the chiller, the optional Model 3930 Lok-to-Clock
electronics module, and the standard Model 3955 electronics module. Refer
to Figure 4-1 or Figure 4-3.
Purge bleed valve—when open, allows more purge gas to flow through the
laser head during initial start-up. It is closed for normal operation.
Water inlet connector—input for the laser rod cooling water from the
chiller.
Water outlet connector—output for the laser rod heated water to the
chiller.
Purge inlet connector—inlet for clean, dry gas from the Model 3910.
LTC PD photodiode BNC connector—sends the reference photodiode sig-
nal to the phase detector circuit in the optional Model 3930 Lok-to-Clock
electronics module. If the latter is not present, a 50 terminator must be
attached.
TO 3930 main signal D-sub connector (25-pin female)—sends and
receives signals to and from the optional Model 3930 Lok-to-Clock elec-
tronics module.
TO 3955 main signal D-sub connector (25-pin male)—receives heater,
power, and control signals from the Model 3955 electronics module.
ML PD photodiode BNC connector—sends the reference photodiode sig-
nal to the acousto-optic modulator (AOM) driver in the Model 3955 elec-
tronics module.
ML mode locker BNC connector—receives the AOM RF signal from the
electronics module.
Control Panel
The laser head control panel indicates whether or not pulses are detected,
and shows the temperature status of the mode locker. For ps systems, it also
provides
GTI temperature status and dispersion control for the GTI (Figure
4-2). This panel is not present on the Model 3941.
PULSING indicator—when green, it indicates either pulses are present or
conditions are correct for pulses.
MODE LOCKER TEMPERATURE indicator—when green, it indicates
mode locker temperature has stabilized; red indicates the opposite. From a
“cold” laser start-up, expect a stabilization time of about 20 minutes.