Alignment
6-3
3. Verify the optics are clean and the proper optics are installed in the
right positions.
Refer to the tables at the end of Chapter 10, “Options and Accesso-
ries,” to match mirror part numbers to the respective mirror positions.
Optic part numbers are written on the barrel of each optic along with a
v-shaped arrow that points to the coated, intracavity surface. Always
close the shutter when replacing optics.
4. Clean the Tsunami windows.
Refer to Chapter 9, “Maintenance: Removing and Cleaning Tsunami
Optics: Brewster Windows.
Figure 6-1: Tsunami fs Configuration.
Figure 6-2: Tsunami ps Configuration.
Output
Brewster
Window
Fast
Photodiode
Beam
Splitter
M
10
M
8
Pr
4
AOM
M
9
Pr
3
Pr
2
Tuning
Slit
M
6
Pr
1
M
7
M
5
P
2
M
3
M
4
PZT (optional)
Ti:sapphire Rod
P
1
M
2
Residual
Pump
Beam Dump
Motorized (optional)
Input
Brewster
Window
Pump
Beam
Model 3955
AOM Driver Electronics
Model 3930
Lok-to-Clock Electronics
Optional
HR
OC
M
1
Output
Brewster
Window
Fast
Photodiode
Beam
Splitter
M
10
AOM
M
5
P
2
M
3
M
4
PZT (optional)
Ti:sapphire Rod
P
1
M
2
Residual
Pump
Beam Dump
Motorized (optional) Birefringent
Filter
Input
Brewster
Window
Pump
Beam
Model 3955
AOM Driver Electronics
Model 3930
Lok-to-Clock Electronics
Optional
HR
OC
M
1

Table of Contents