I. Visual inspection OK Not
ok
Description
of Defect
What? How?
Patient lifting
pole, adapter
sleeves, grab
handle with strap
No damage or deformation
Manufacturer's recommendation: Replace grab handle after 5
years
Bed frame: Mat-
tress base, chas-
sis
No damage, deformation, no split welded
seams
Castors No damage
Mattress base No damage or deformation
Safety sides No damage, cracks or deformation
Connecting ele-
ments (screws,
bolts, nuts, safe-
ty caps)
Secure fixing, missing parts
Wearing parts,
joints
No damage or severe wear
Plastic parts, var-
ious
Free of sharp-edged damage/breakage
II. Electrical Measurement according to DIN EN 62353 (VDE 0751-1)
Limit val-
ue
Measured
value
Earth conductor resistance measuring
point: PE pin on the bed, head end,
(recommended measuring current for
beds cleaned in automatic washing
systems ≥ 2 A)
0.3 Ω ____ Ω
Device leakage current type: B, direct
(place bed with conductive castors in a
way that it is insulated) or difference.
0.5 mA ____ mA
Appendix