Level -
Type of trigger, enabling the event recording to start
for the Recording Time under the condition: value of
the RMS result (Source) integrated by the 0,5 ms
period is lower than the threshold level (Level). In
other cases, the recording doesn’t start, except
when the recording is already in progress during the
Recording Time.
Manual
5.5.4
Gradient +
Type of trigger, enabling the event recording to start
for the Recording Time under the condition: value of
the RMS result (Source) integrated by the 0,5 ms
period is greater than the threshold level (Level) and
the gradient of this Source is greater than the
threshold level (Gradient). In other cases, the
recording doesn’t start, except when the recording
is already in progress during the Recording Time.
Manual
5.5.4
Trigger manual
Type of trigger, enabling manual triggering of the
recording start after pressing simultaneously and
keys during the measurement.
Manual
5.5.4
Integr. Period
Type of trigger, enabling the signal recording every
time the measurement starts, and the recording will
last minimum Recording Time. If the triggering
condition appears during recording (when
Integration Period is shorter than Recording Time)
from this moment the recording will be continued for
the next Recording Time and so on.
Manual
5.5.4
Filter
Weighting filter used during event recording: Z, A,
C, B or LF.
Manual
5.5.4
Sampling
Sampling frequency of the event recording: 12 kHz.
Manual
5.5.4
Source
Result that is compared with the threshold level
(Level) for triggering LEQ measured in the first
profile: Leq(1).
Manual
5.5.4