SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP
INTRODUCTION
1.2. TECHNICAL INFORMATION
Parameter Whole blood mode Pre-diluted mode
Repeatability
(With 95% reliability limit)
WBC (≥ 4.0× 10
3
/μL)
RBC (≥ 3.00 × 10
6
/μL)
HGB
HCT
MCV
MCH
MCHC
PLT (≥ 100 × 10
3
/μL)
LY M #
MXD# (≥ 1.0 × 10
3
/μL)
NEUT#
LY M%
MXD% (≥ 12%)
NEUT%
RDW-SD or RDW-CV
PDW
MPV
P-L CR
3.5% or less
2.0% or less
1.5% or less
2.0% or less
2.0% or less
2.0% or less
2.0% or less
6.0% or less
15.0% or less
30.0% or less
15.0% or less
15.0% or less
30.0% or less
15.0% or less
4.0% or less
12.0% or less
5.0% or less
20.0% or less
6.0% or less
3.0% or less
2.5% or less
3.0% or less
3.0% or less
3.0% or less
3.0% or less
9.0% or less
—
—
—
—
—
—
—
—
—
—
Accuracy
WBC
RBC
PLT
± 3% or ± 0.2 × 10
3
/μL or less
± 2% or ± 0.03 × 10
6
/μL or less
± 5% or ± 10 × 10
3
/μL or less
± 5% or ± 0.3 ×10
3
/μL or less
± 3% or ± 0.05 ×10
6
/μL or less
± 8% or ± 15 × 10
3
/μL or less
(Excluding the dilution error at sample preparation.)
Linearity in whole blood mode
(applies, when RBC < 7.00 ×10
6
/μL)
WBC
1.0 – 9.9 (×10
3
/μL)
10.0 – 99.9 (×10
3
/μL)
± 0.3 (×10
3
/μL) or less
± 3% or less
RBC
0.3 – 0.99 (×10
6
/μL)
1.00 – 7.00 (×10
6
/μL)
± 0.03 (×10
6
/μL) or less
± 3% or less
HGB
0.1 – 10.0 (g/dL)
10.0 – 25.0 (g/dL)
± 0.2 (g/dL) or less
± 2% or less
HCT
10.0 – 33.3 (HCT%)
33.3 – 60.0 (HCT%)
± 1.0 (HCT%) or less
± 3% or less
PLT
10 – 199 (×10
3
/μL)
200 – 999 (×10
3
/μL)
± 10 (×10
3
/μL) or less
± 5% or less
1.2.1. Performance characteristic - Specifications
INTRODUCTION
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP