CLEANING & MAINTENANCE
INTRODUCTION
pocH-100
i
Revised February 2011
4.1.10. Drain reagents
Performbeforemoving/shippingtheinstrument.
Remove spout kits from diluent and lysing
reagent.*
As needed
Press"Menu".
Press"Maint.".
Press"DeprimeSeq.".
Press"Execute".
Processbegins.
SwitchO F F, drainandcleanwastebottle.
Always wear protective garments and gloves
whenhandlingthewasteuid.Afterworkiscom-
pleted,washhandswithdisinfectant.
Biological risks
Remove spout kits from diluent
and lysing reagent.
Press "Menu".
Press "Maint.".
Press "Deprime Seq.".
Press "Execute".
Process begins.
Switch OFF .
The screen of process completed.
•Placetheremovedcontainerspoutkitsontopof
acleanclothorinsideaplasticbag,andshield
fromdust.
•Theinstrumentwillbeprimedautomaticallythe
nexttimeitisstartedup.
*ThisisimageofthepocH-pack65.ThepocH-pack
65XLdoesnothavewastecontainer.
Note
Thescreenofprocesscompleted.