EasyManua.ls Logo

SYSMEX pocH-100i - 5.2. Calibration Procedures; 5.2.1. Calibrate the Instrument

SYSMEX pocH-100i
105 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1234 678
5
5 - 11
pocH-100
i
Revised June 2008
SETTINGS & CALIBRATION
INTRODUCTION
5.2. CALIBRATION
5.2.1. Calibrate the instrument
If quality control shows repeated deviations in the same direction.
If a major component of the instrument has been replaced.
HGB Determination of hemoglobin concentration (DIN 58931)
HCT Determination of the concentration of blood corpuscles in blood (DIN 58933)
Recommended measuring methods:
Establishing reference values
The HGB and/ or HCT values are corrected by a calibration value.
Calibration does not need to be performed at
specific intervals. Follow your internal laboratory
regulations for performing calibration.
Each sample should be analyzed at least three
times.
EIGHTCHECK is not suitable for calibration pur-
poses, but for quality control instead.
Important
Samples used for calibration
Use 5 or more samples of fresh blood meeting the following conditions:
• Blood from healthy person who is not taking any medication
Blood added with an appropriate amount of anticoagulant
Per-sample whole blood volume to exceed 2 mL
• HGB value to exceed 10.0 g/dL
HCT value to be within 35.5% and 55.5%

Table of Contents

Related product manuals