2.1. ANALYSIS
2.1.6. Analysis in pre-diluted blood (PD) mode
Check blood which has been collected in skin puncture diluted at the ratio 1:26. Aspirated sample volume~200 μL. Use of EDTA anticoagulant.
• Required sample volume: >20 μL for the ø 13mm tube, > 500 μL for the micro tube.
* Steps 11 and 12 are not necessary in case of open pipette.
Press "PD".
Open sample position.
Insert the correct adapter.
Insert an empty micro tube (cap removed)
and close door.
Press "Execute".
500 μL of diluent are dispensed into the tube.
After dispensing, press "OK" and remove the
micro tube.
Dispense 20 μL of whole blood. Cap micro
tube and mix gently.
Press "Sample ID", enter the ID manually or
by barcode reader and press "Ent.".
Press "PD".
Open sample position.
Insert adapter.
Insert an empty micro tube (cap
removed) and close the door.
Press "Execute".
After dispensing, press "OK" and
remove the micro tube.
Dispense 20 μL of whole blood.
Press "PD".
Press "Sample ID". Enter the ID
and press "Ent.".
Press "Dispense".
Press "Dispense".
Press the -button or press "Operator".
Press the -button or press "Operator".
Press "PD".