CLEANING & MAINTENANCE
INTRODUCTION
4.1.6. Clean the sample tube adapter
SwitchO F F.
Removeadapterandcleanit.
- Clean with distilled water
or
- Clean with CELLCLEAN diluted 10 times or
more.
Rinse the adapter with distilled water after-
wards
As needed
Ifsamplehasspiltintotheadapter.
Switch OFF.
Remove adapter and clean it.
Always wear protective garments and
gloveswhenhandlingthe waste fluid.
After workiscompleted,washhands
withdisinfectant.
Biological risks
4.1.7. Perform auto rinse
Cleantubingsanddrainwaste.
Evenifthemessage"AutoRinse"isnotdisplayed,autorinsecanbeexecutedbypressing
"Menu","Maint."andthen"AutoRinse".
As needed
Process starts.
Background check result will ap-
pe ar.
Press"Execute".
Process starts. Afterwards "Background
check"willbeperformedautomatically.
Press "Execute".
Press"Result".Backgroundcheckresult will
appear.