CLEANING & MAINTENANCE
INTRODUCTION
pocH-100
i
Revised February 2011
4.1. CLEANING
4.1.13. Replace reagent
As needed
Insert the container spout kit into the new
pocH-pack65.Tightenthecap.*
Press"Execute".
Enter reagent barcode by barcode reader
andproceedtostep7orpress"Manual".
Enterbarcodeofthenewreagentbottle.
Insert the container spout kit
into the new pocH-pack 65.
Press "Execute".
Enter barcode of the new rea-
gent bottle
Press"OK".
Press "OK".
Press"Execute".
Afterreplacemenuprocess,displayshows
startscreen.
Press "Execute".
After replace menu process, dis-
play shows start screen.
The screen displays "Replace
pocH-pack65".
Thescreendisplays"ReplacepocH-pack65".
Enter reagent barcode by bar-
code reader and proceed to step
7 or press "Manual".
•Donot to touch thepipe withyourhands, or
to allow dust or other substances to adhere to
them. Ifsuch substancesadhere to the pipe,
wash the substance off with reagent first before
attachingthecontainerspoutkit.
Failuretodosomayleadtoerroneousresults.
•Becarefultopreventthereagentfromspilling.
Ifthereagentspills,immediatelywipeitoffwith
adampcloth.
*ThisisimageofthepocH-pack65.ThepocH-pack
65XLdoesnothavewastecontainer.
Important
•Usereagentat15–30°C.
Leave reagents at room temperature for at least
24hoursorlongerbeforeusing.If arefrigerated
reagentisused,a correctanalysisresultwillnot
beobtained.
•Fordetailedinformationaboutstorage&shelf
life,pleasesee"Thereagents"page1-4.
Incorrect analysis results
•Wasteuidmaydiscolorsomematerials.
•Ifthewasteuidisspilt,wipeoffwiththedamp
clothimmediately.
•Cleanspillsaccordingtoyourpoliciesandproce-
dures.
Caution
Whenhandling(disposing)thewasteuid,dispose
it appropriately in accordance to local laws and
regulations, whichconsiders forthe disposalof
medicalwasteandinfectiouswaste.
Biological risks