ANALYSIS
INTRODUCTION
Incorrect analysis results
All performance data cited in this manual were
generated using specimens in EDTA anticoagulant.
Results may differ using other anticoagulants.
Note
The sample ID can be set up to 15 characters. If a sam-
ple‘s ID is set to „ 0 “, the result for this sample will not
be stored.
• When using micro tubes, place them in the cor-
rect adapter and make sure the cap is removed.
• Only use pierceable rubber caps. To avoid intro-
ducing rubber fragments from the cap affecting
the measurements, use each rubber cap only
once.
• Using resin caps may damage the piercer! Do
not analyze with resin capped tubes!
*
1.
Do not open the sample position while aspirat-
ing!
Important