V/D
S
ERIES
E
LECTRICAL ASSEMBLIES
Revision: 2.1 4-23
4.2.5 Relay board for motor control
Fig. 23: Relay board for motor control on the D-23
1 JP3 Gear motor
2 JP1 Power supply
3 JP2 Digital input for motor control
4.2.6 Motor control board for radial fan
• See chapter 4.1.6 oben.
4.2.7 Fuses on the back of the device
The following safety elements and fuses (see chapter 2.2 above) are
located on the back of the device:
− Safety temperature limiter (STL)
The safety temperature limiter monitors the heating element and triggers if
the temperature exceeds 170 °C. The heater contactor K1 switches off
and cuts the power to the solid state relays (SSR).
Rose-D
Board
JP6
N
L
Res.High elec.(E6)
HE AT ER
ELEMEN T
2900 W
PANL1002-V2
Rose-A
Board
1
2
3
4
5
6
Chamb.high Elec.(E3)
15VDC
4-20mA
POW5
POW4
POW3
POW1
POW2
PB
PA
PA
VCC
PB
GND
WATERPUMP
PULL SOLENOID
VAC.PUMP
VCC
GND
AC1
L
GND
N
JP1
1
2
3
4
JP18
1
2
3
Y12
VAC.BREAK.
12
JP1
JP7
1
2
JP9
JP3
1
2
3
PULL S olenoid D oor open Y27
12
STEAM CH.
JP5
1
2
3
4
5
6
7
FLOW SW.VAC.
12
JP4
1
2
3
4
5
6
Y16
FASTEXH.TOP.
12
12
JP11
1
2
JP15
1
2
3
4
5
6
7
JP5
1
2
3
MainSwitch
JP2
1
2
3
FLOW SW.WAT.
12
JP7
HEADER 3
1
2
3
JP6
1
2
3
4
5
DOOR LOCKED
12
JP12
1
2
5
6
7
8
JP13
1
2
3
4
5
6
7
Y7
FASTEXH.BOT.
12
H2
JP10
1
2
3
4
5
6
7
8
9
10
JP17
1
2
3
4
5
6
JP12
JP8
1
2
3
4
5
JP14
JP13
1
2
3
JP2
JP14
HEADER 2
1
2
JP2
1
2
3
4
JP16
1
2
3
4
5
6
JP9
1
2
3
4
DOOR CLOSED
12
DOOR LOCK
MOTOR 12 DC
1
2
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
4
3
1
2
DOOR UNLOCKED
1
2
1
2
3
12
FAN
(N24 VAC)
(N24 VAC)
AC1 (24 VAC)
AC2
N
N
GND
13 VDC
5 VDC
GND
5 VDC
13 VDC
GND
5 VDC
Main Filter
5V
DATA
5V
GND
GND
5V
DATA
CHAMBER
PRESS
+
-
Huba Control
1
2
3
13 VDC
GND
5 VDC
OUT
JP2
1
2
JP3
JP1
1
2
Rel-Mot Board
JP16
Fuse 5x20 - 1,5A
Fuse 5x20 - 1,5A
Fuse 5x20 - 3,15A
Fuse
5x20 - 3
,15A
Rose-
P Board
F1 F2
F3
F4
“N”
Terminal
Cut off (STB)
12
Y2
H1
(AC1)
(AC4)
5V
GND
F 1 - Free (3,15A)
F 2 - Vacuumpump fuse (3,15A)
F 3 - Water pump fuse (1,5A)
F 4 - PULL Solenoide fuse (1,5A)
F 5 - Primary fuse (800mA)
F 6 - Secondary fuse (5A)
F 7 -
F 8 -
F 9 -
F11- Resettable motor control fuse (3,5A)
Systec DX23
A3
Size:
Date: 28.01.2010
Drawing No:
Drawing name:
Blatt:
1 of 1
Wiring diagram
Systec DX23
F10 -
Heating element overview
DX23 = 2900W (230V)
24VAC-1
24VAC-2
115VAC
115VAC
12VAC
Toroidaltransformer
(230V to 24V)
12VAC
F6
12VAC
F5
JP15
1
2
24
VAC
5 VDC
GND
PRINTER
DATA
MODUL
CN2
1
2
3
4
CN5
1
2
3
4
1
2
3
4
5
6
7
8
9
10
5V
GND
GND
5V
RXD
TXD
DTR
GND
CIRC. PUMP
A2
1
K1
2
A1
RADIAL
VENTILATOR
MOTHERBOARD
24 VAC
M
-
Radial Fan
24V DC
+
X 1.9
X 1.8
X 1.7
X 1.6
X 1.5
X 1.4
X 1.3
X 1.2
X 1.1
! Modified Version !
Ground isn’t connected
F11
VACUUM
12
Y3
WATER COIL
AIRCOOLING
M9
OPTIONAL
(AC3)
VAC PUMP M2
1 2
VAC.WATER
ValveY15
Gen.High elec.(E1)
Gen.Low elec.(E2)
Chamber Temp
RED
WHITE
WHITE
Cond. Temp
WHITE
RED
WHITE
PT100 Temp2
WHITE
RED
WHITE
MPX5700
1
2
3
GEN PRESS
GND
VCC
Y4
COOLING DRAIN
12
Y11
FLUSH COIL
12
Y5
COMP.AIR CH.
12
Y9
12
CONDENCE
Y1
STEAMCOIL.
12
1000?
1
2
Waterpump
M1
M8
M4
Generator
PRESS
RED
WHITE
WHITE
PT-100
PT-100
PT-100
Drain Temp
M
(AC5)
12
Y17a
Water to Res.
1
2
3
4
Elec. aquastop (E5)
Elec. plate(E7)
Aquastopconnector
FLOA T SWITCH RES.
US B
Rose-Com
Board
SD
Card
1
2
3
4
5
P2
JP1
1
2
3
4
5
DB9
5
9
4
8
3
7
2
6
1
6
7
GND
GND
5V
RSO
RSI
GND
RSI
RSO
GND
RSI
RSO
RSI
RSO
GND
GND
5V
JP3
Fig. 24: Fuse block: Position in the circuit diagram