EasyManua.ls Logo

Systec DB-45 - Testing the Sterilisation Chamber for Leaks - Vacuum Test

Default Icon
219 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
V/D
S
ERIES
M
ECHANICAL ASSEMBLIES
3-76 Revision: 2.1
Reattach the pump cover.
Attaching the pump cover:
Ensure that the O-ring is seated properly in the groove.
Proceed as follows to remove scaling:
Prepare a container with a scale remover solution (scale remover and
warm water at approx. 50 °C).
Dip the two hoses into the solution and connect to the pump inlet and
outlet.
Fix the dirt trap (sieve or filter) to the hose end used for solution intake.
Let the pump run for about 30 minutes.
Repeat the process several times.
Testing the sterilisation chamber for leaks – vacuum test
Have available Leak detection spray (Systec item no. 11922)
Vacuum test
Only carry out a vacuum test when the sterilisation chamber is
cold.
During the vacuum test program, you can check PumpFlowSw
to see that it switches on and off in the Service ->
Digital inputs menu.
A leak test is made as follows:
1. Either use program 13 (“Vacuum test”), or
2. Carry out a vacuum test on the sterilisation chamber as follows:
Evacuate the sterilisation chamber.
Close all valves.
Check the pressure.
If the pressure is constant, the autoclave has passed the vacuum test.
If the pressure is not constant, there is a leak. Common sources of leaks
are as follows:
Temperature sensor in the sterilisation chamber (TS1)
Sterile air filter
Slanted filter
Search for the leak as follows:
Admit cold compressed air into the sterilisation chamber, for example
via the validation port or the support air connection, when available
(use the C o mpAirChamb output in the Service ->
M a nual output set menu).

Table of Contents

Related product manuals