Error message: Low vacuum
YesNo
After approx. 40 min. (parameter 68, “PulsTimeOut”),
the vacuum was not reached in the pulses before
the sterilisation phase
Vacuum pump running?
- Check the fuse
(
see chapter 3.11,
“Vacuum unit”
)
- Vacuum pump worn
- Water usage of the pump not suitable
(
see chapter 3.11, “Vacuum unit”)
- Leak on the sterilisation chamber
(
see chapter 3.10.4, “Checking the
compressed air system for leaks”
)
Vacuum pump is running,
but does not reach the set value
(parameters 7/10/13, PlsVac 1/2/3)