EasyManua.ls Logo

Systec DB-45 - Principle; Closing

Default Icon
219 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
V/D
S
ERIES
M
ECHANICAL ASSEMBLIES
Revision: 2.1 3-39
Measure the distance from the outer edge of the door to the lip. If the
distance is less than 17 mm, you must replace the door seal.
Pull out the door seal.
Insert a new door seal.
3.3.3 D-23
Principle
The devices are equipped with an automatic door lock. By locking the
door, two lock catches mesh with each other: The door is now in the
closed position.
After starting the program, the lock catches are moved into the locked end
position by a gear motor. The pressure tightness is only guaranteed in the
locked end position.
The controller then checks the correct procedure using a series of inputs
and outputs, thus ensuring safe operation of the device.
Closing
Door Open is shown in the display when the door is open.
Close the door so that the lock catches mesh with each other. The
“Door Closedreed contact is closed and the Door Open message no
longer appears: The door is now in the closed position.
Starting the program triggers a chain of events to lock the door securely.
The gear motor is started and moves both lock catches via a spindle into
the locked end position, which is defined by the “Door Locked” switch.

Table of Contents