V/D
S
ERIES
T
ROUBLESHOOTING
7-2 Revision: 2.1
7.1 Reset
Only carry out a reset at atmospheric chamber pressure
The reset function may only be used when at atmospheric
chamber pressure (max. 10 kPa above atmospheric
pressure).
If the program sequence falters, the controller software can be reset:
− The control unit is returned to its basic state.
− Program 1 is selected.
− In the event of excess pressure, the atmosphere valve is opened.
− If there is a vacuum, the vacuum breaker is activated.
− The cover can be opened once atmospheric conditions are reached,
regardless of the temperature measured inside the device.
is a risk of explosion if there is fluid in the device
at temperatures above 80 °C!
Do not perform a reset if there are hot fluids in the device – let
the device cool down gradually first.
If there is excess pressure in the device, go to Menu >
Service > Manual output > Set before resetting
and activate the fast exhaust valve to let out the excess
pressure.
Procedure:
• Turn off the device using the main switch.
• Press and hold the right display button.
• Turn on the device using the main switch.
• The display first shows the message Systec Autoclaves, followed
by Reset!!!.
• Release the right display button within one second.