V/D
S
ERIES
C
OMMISSIONING
Revision: 2.1 2-19
2.10.6 Changing the print intervals
If the device is equipped with a printer, the print interval times
(parameters 55/56) can be changed using the service code (level 4).
This function allows you to reduce paper consumption and only record
important phases in detail.
The default settings are 30 seconds for the sterilisation phase and
300 seconds for all other phases.
More information on printer functions and operation can be found in the
operating manual (see chapter 6, “Options”).
2.11 Reference
• Please complete the following documents:
− Service report
− Logbook
− Installation log
− Commissioning log
− Instruction log
Have the instruction log signed by the operating personnel.