7.3.3 Error message: High Gen.Press
Error message: High Gen.Press
The pressure in the steam generator is
higher than permitted according to
parameter 69 (“GenMaxPress”).
- Pressure sensor on steam generator
defective
- Check the functionality of the steam
generator – there is a possibility that the
water inlet has not been stopped and the
steam generator is then filled with the line
pressure from the demineralised water line
(
see chapter 3.8.1, “Steam generator”)
- Check the factors for the steam generator
(
parameters 36 and 37) with regard to the
sterilisation temperature of the program.
(consult Systec, when necessary)