V/D
S
ERIES
S
ERVICE SOFTWARE
Revision: 2.1 6-15
6.2.6 Download Analog (only for D-23)
On type D-23 autoclaves, you can overwrite the firmware on the analogue
circuit board. An error message appears if this function is attempted on a
different type of autoclave:
Fig. 13: Error message “Wrong download module. Action aborted”
6.2.7 Calibration
You can modify the calibration of the analogue device inputs.
The measurement range during the calibration must
correspond to the conditions of use!
Temperature sensors do not measure in a completely linear
manner. This means that the calibration is more exact the
closer the calibration measurement comes to the actual
subsequent use.
Systec references:
− Pressure between 100 and 300 kPa
− Temperature between 80 °C and 130 °C
− Drain temperature between 20 °C and 80 °C