Checking the water feed pump (M1)
Check that the impeller works correctly, and for unusual noises and
correct conveying.
The impeller is available as a separate spare part.
Check the pump capacity (only for external water supply):
If the external water supply comes from a water container (e.g. a canister),
then the pump may suck in air.
When the water feed pump (M1) has sucked in air, then water is no longer
conveyed: The pump switches off as the flow monitor no longer registers
any water flow. The NoDeminWater error message appears.
Air extraction:
Unclip the hose between the water feed pump (M1) and no-return valve
(R1) at the no-return valve, then feed it into a container.
Switch the equipment on at the main switch.
Let the air bleed out of the water feed pump until water flows.
Switch off the device and reattach the hose.
Checking the functionality of the flow monitor
V-40 – 55 V-65 – 150 X E B
 
D-23 D-45 D-65 – 200
X E B

Check GenFlowSw: Whilst the steam generator is being filled by the
water feed pump, you can check GenFlowSw to see that it switches on
and off in the Service -> Digital inputs menu.
Pay attention to the level of water in the container!
Ensure that the container is always filled with enough water.