ECHANICAL ASSEMBLIES
3-42 Revision: 2.1
Setting the safety ramp and door switch
The safety ramp and door switch must be set together.
Close the door.
1
2
3
Fig. 19: Loosening the screws on the limit switch
Loosen the screws on the limit switch (1).
Trigger the Close door digital output in the Service -> Output
set display until the vacuum lip of the door seal lays flat on the collar
of the sterilisation chamber.
Position the “Door Locked” limit switch.
The position of the support plate (2) can be set using the slotted holes (3),
when necessary.
Tighten the limit switch: The lock catch of the door must now protrude
5 mm over the safety ramp.
When necessary, correct the position of the safety ramp and secure it
in place using the required locking plates.
Move the motor to the unlocked position using the Open door digital
output: The lock catch must now be positioned 2 mm away from the
safety ramp.
Position the “Door Unlock” limit switch and screw it tight.