Setting the operating modes
The following operating modes can be set using parameter 93
(WaterTime):
WaterTime = 0
The demineralised water is pumped into the sterilisation chamber via a
water feed pump when the program is started and during the sterilisation
phase (optional).
The demineralised water is drained out of the sterilisation chamber after
the end of the program.
WaterTime = 1
The sterilisation chamber must be filled manually with demineralised water
before the program is started.
The demineralised water is drained out of the sterilisation chamber after
the end of the program.
WaterTime = 2
The sterilisation chamber must be filled manually with demineralised water
before the program is started.
After the end of the program, the demineralised water remains in the
sterilisation chamber when the end temperature (parameter 5, EndTemp)
measures ≤ 80 °C. Otherwise, the demineralised water is also drained out
of the sterilisation chamber.
WaterTime > 2
This is the run-on time in seconds for the intake of demineralised water
after the chamber electrode (E3) has detected water.
The demineralised water is drained out of the sterilisation chamber after
the end of the program.
Water feed pump
V-40 – 55 V-65 – 150 X E B
D-23 D-45 D-65 – 200
X E B
The water feed pump (M1) is a vibration pump (see Fig. 32: oben, pos. 3)
with impeller.