V/D
S
ERIES
S
ERVICE SOFTWARE
Revision: 2.1 6-23
Set
button
The displayed value overwrites the software version
number stored in the device.
Load number
The batch number of the device is displayed.
You can change the value manually.
Get
button
The value stored in the device overwrites the displayed
batch number.
Set
button
The displayed value overwrites the software batch
number stored in the device.
Replace filter
The filter counter value (the number of sterilisation
cycles with this filter) is displayed. You can change the
value manually.
Get
button
The value stored in the device overwrites the displayed
number of sterilisation cycles with this filter.
Set
button
The displayed value overwrites the number of
sterilisation cycles with this filter stored in this device.
Check Hardware
button
The function of all memory devices attached to the
internal bus is tested.
If the hardware check fails, this may be due to the following:
EEPROM of the analogue circuit board
EEPROM of the extension board (if present)
Ribbon cable between the two boards
Errors
Select “Errors” to see a list to the ten most recent errors. You can save
this as an rtf file and print it out.