ECHANICAL ASSEMBLIES
3-56 Revision: 2.1
7
6
8
9
1
3
4
2
5
Fig. 31: Filling the closed cooling circuit
1 3-way ball valve (BV3)
2 Pump (M6)
3 Sterilisation chamber
4 Cooling coil
5 Canister with water
6 Cooling panel (HE)
7 Manometer (P3)
8 Surge tank (PV3)
9 2-way ball valve (BV2)
Connect the free connections from the 3-way and 2-way ball valves
(BV3 and BV2) with the canister using the appropriate hoses (i.e.
immerse the loose hose ends in the water).
Open the 3-way and 2-way ball valves and control the pump in the
Service -> Manual output set -> CircPUMP menu until
the system is filled.
Stop the pump when gas formation is no longer visible. Close the
valves and disconnect the hoses.
Connect the 2-way ball valve to the water supply and slowly open the
valve until the required system pressure is reached (approx. 1.5 bar
when cold, see the manometer).
Close the valve and remove the water hose.
After some time, observe the manometer reading and check whether
the pressure has fallen (this may indicate a leak).