Various menu settings
Icon name Function Initial setting value
Reference
page
Alarm Setup List
Displays the settings of each menu by
list.
The list of admin settings set under the
each menu of the admin setup.
page 92
BSA
Inputs the Patient’s height and weight
to calculate the BSA value and
perfusion index value.
Height: ---.-
Weight: ---.-
page 93
Count Down
Sets the count down timer. 0:00:00 page 94
History
Displays the trend chart and the event
list. The history can be exported to a
USB memory drive as log data.
page 95
Admin Setup
Opens the admin setup menu.
page 99
The threshold value (setup range: 0.0 to 9.9 LPM (L/min)) of the Flow Alarm can be set.
1. Tap the Flow Alarm icon. (Figure 3-1)
The Flow Alarm screen opens.
2. Confirm that [Alarm] of the upper limit is selected.
(Figure 3-2)
Figure 3-1
3. Tap or to input the upper limit value you want to set.
(Figure 3-3)
“
(red)” indicates the threshold value of the High Flow Alarm in accordance
with an input flow value.
Figure 3-2
Figure 3-3
84
Setting Flow Alarms