EasyManuals Logo

Terumo CAPIOX EBS Circuit with X coating User Manual

Default Icon
177 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #149 background imageLoading...
Page #149 background image
In order to ensure the safe use of this system, make sure it is cleaned after use and stored in an appropriate environment. If the
system is used while it is dirty, the system may not operate properly or may cause a malfunction.
Cleaning
Cautions
Since this system does not have an airtight structure, it should not be used or stored in an active gas environment
(including sterilizer gas), nebulizer-sprayed environment, high-humidity environment, etc. Taking the system to these
environments leads to the electronic components inside the system being affected and there may be subsequent damage
and time degradation which will cause failure of this system.
Clean the system before and after using. When disinfecting, do not use a sterilizer. Use a soft cloth (damp with
disinfectants) to wipe the system, and wipe off the disinfectants with a soft cloth (damp with cold/lukewarm water), and then
thoroughly wipe off any moisture with a dry soft cloth.
If this system is tainted with blood, wipe the system immediately with a damp (not wet) soft cloth. Be sure to wear gloves to
prevent infection.
When liquids (bloods or drug solution, etc.) come into contact with the Drive Motor, immediately wipe them off and clean
the device thoroughly. If coagulated liquids (bloods or drug solution, etc.) are adhered to the pump receptacle of the Drive
Motor, the slide hook may not move.
Turn the system off and disconnect the AC power cable and cables (the Drive Motor, the ABD/Flow Sensor, the
Temperature Sensor Cable, the Pressure Sensor Cable, the Fast Clamp, the LAN cable, the CDI Communication Cable)
before cleaning. Otherwise the system may fail or an electrical shock may occur.
Do not use an electric dryer to dry this system. This system may be damaged.
Do not wash this system with running water or while immersed in water as this may cause damage or failure of the system.
This system does not have a waterproof structure.
Do not wipe with organic solvent such as alcohol or thinner, or with povidone-iodine. Using organic solvent or any
disinfectants other than those permitted for use may result in damage to or failure of this system.
Do not subject this system to ethylene oxide gas (EOG) sterilization or high pressure steam sterilization, or immerse this
system in disinfectant solution.
Since drug solution may cause a short circuit, ensure that the connecting sections (the Drive Motor, the ABD/Flow Sensor,
the Pressure Sensor Cable, the Temperature Sensor, the Fast Clamp, the LAN cable and the CDI Communication Cable)
are not wet when connecting plugs to connectors. If moisture is present, ensure that the power is turned off and the AC
power cable is removed from both this system and grounded AC power source, and then thoroughly wipe with a dry
soft cloth. Since this system does not have a waterproof structure, drug solution and moisture may affect the electrical
components inside, and cause a malfunction.
120
After Use

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Terumo CAPIOX EBS Circuit with X coating and is the answer not in the manual?

Terumo CAPIOX EBS Circuit with X coating Specifications

General IconGeneral
BrandTerumo
ModelCAPIOX EBS Circuit with X coating
CategoryMedical Equipment
LanguageEnglish

Related product manuals