EasyManuals Logo

Terumo CAPIOX EBS Circuit with X coating User Manual

Default Icon
177 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #120 background imageLoading...
Page #120 background image
Various menu settings
2. After tapping [On] or [Off], tap [OK] to complete the setting.
(Figure 3-30)
Figure 3-30
• If you do not tap [OK], the changes to the setting will not take effect.
• The air bubble detection sensitivity can be adjusted only through the admin setup.
<Indication of the ABD status icon>
The icon indication changes in accordance with the status of the air bubble detector function.
For details, refer to “LCD Touch Panel Display (When Operating Normally)” (page 9).
Response behaviours (Message Only/Pump Coast/Clamp+Pump stop) to the Over/Low Pressure Alarm, Air Bubble Detected
Alarm and Back Flow Alarm can be set.
1. Tap the Safety Connection icon. (Figure 3-31)
The Safety Connection screen opens.
2. Tap the response behaviour you want to set for each alarm.
(Figure 3-32)
[Message Only]:
Displays a message when the alarm is detected.
[Pump Coast]:
Displays a message when the alarm is detected and changes the motor
speed to the coast speed.
[Clamp+Pump stop]:
Displays a message when the alarm is detected, closes the
extracorporeal circulation circuit line with the Fast Clamp and stops the
Drive Motor.
3. Tap [OK] to complete the setting. (Figure 3-33)
Figure 3-31
Figure 3-32
Figure 3-33
91
Notes
Setting the Safety Connection

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Terumo CAPIOX EBS Circuit with X coating and is the answer not in the manual?

Terumo CAPIOX EBS Circuit with X coating Specifications

General IconGeneral
BrandTerumo
ModelCAPIOX EBS Circuit with X coating
CategoryMedical Equipment
LanguageEnglish

Related product manuals