Operation Procedure
1. Press the power button. (Figure 1-40)
2. The LCD touch panel display is lit, and the Controller is
activated.
After the start screen is displayed with an activation sound, the main screen
appears. (Figure 1-41)
Figure 1-40
Figure 1-41
47
Warnings
• Use the AC power source as the primary power source. Battery is used as an auxiliary power source when the AC
power source cannot be used sufficiently, such as during transportation, electric failure, etc. If the battery is depleted
and charging is insufficient, the Controller will not be able to operate by the battery during an emergency such as electric
failure.
• When the Controller signals the Serious System Error (device alarm (high priority)), attach the centrifugal pump with the
Drive Motor connected to the Backup Controller or the Hand Crank to maintain circulation. Otherwise, the pump may stop
and may cause injury to the patient.
Turning the Power On
Power button