Maintenance and Inspections
Serial number: Date: / / Person in charge:
Check items
Inspection before each use
Inspection of the
external appearance
(Before each use)
Are there any significant scratches or damage in the appearance?
Are there any scratches or pinch marks on the outer covering of the cables?
Are there any foreign items or damage in the connectors?
When turning the motor speed knob, does it turn smoothly?
Pass
Pass
Pass
Pass
Fail
Fail
Fail
Fail
Inspection of start-up
sequence (Before
each use)
Do the following buttons and indicators in the control panel flash three times?
Auto priming button
• Coast release button
Key lock button
• Flow indicator
Motor speed indicator
Does the status lamp flash from green to yellow to red?
Does the motor speed indicator lamp turn from green to orange in a clockwise
direction?
Does the battery indicator lamp flash twice repeatedly in the following sequence?
(None, none, red) (None, green, green) (Green, green, green)
When the start screen is displayed, is an activation sound heard?
Does the main screen appear?
<Confirmation items after the main screen is displayed>
Is the Serious System Error or the System Error issued?
Is the AC power lamp lit and is the AC power indicator
?
Is there more than 70% battery charge level?
Pass Fail
Pass
Fail
Pass Fail
Pass Fail
Pass
Fail
Pass Fail
Pass
Fail
Pass Fail
Pass Fail
Inspection of the
battery-powered
operation function
(Before each use)
<When the AC power cable is unplugged>
Is the Battery-powered operation started Alarm issued?
Does the AC power lamp turn off?
Does the AC power indicator change to
?
Pass
Pass
Pass
Fail
Fail
Fail
<When the AC power cable is plugged in>
Does the On Battery Alarm stop? Pass Fail
Does the AC power lamp light up? Pass Fail
Does the AC power indicator change to
?
Pass Fail
Inspection of the Drive
Motor Disconnected
Alarm and the
ABD/Flow Sensor
Disconnected Alarm
(Before each use)
<When the plug of the Drive Motor is disconnected and the power is turned on>
Is “Drive Motor Disconnected” displayed on the message area and does an alarm
sound?
Does the alarm mute when tapping
?
<When the plug of the ABD/Flow Sensor is disconnected and the power is turned on>
Is “ABD/Flow Sensor Disconnected” displayed on the message area and does an
alarm sound?
Does the alarm mute when tapping
?
Pass
Pass
Fail
Fail
Pass Fail
Pass Fail
129
SP-200 Maintenance/Inspection Check List

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Terumo CAPIOX EBS Circuit with X coating and is the answer not in the manual?

Terumo CAPIOX EBS Circuit with X coating Specifications

General IconGeneral
BrandTerumo
ModelCAPIOX EBS Circuit with X coating
CategoryMedical Equipment
LanguageEnglish

Related product manuals