Operation Procedure
4. After confirming the message displayed and expose the
pressure sensor to the air, tap [OK]. (Figure 1-50)
Calibration of the pressure sensor is started.
Figure 1-50
When calibration has been finished normally, “0” will be displayed at the side
of [CAL]. (Figure 1-51)
When any number other than “0” or “----” is displayed, perform calibration
again.
Figure 1-51
When using the Fast Clamp, confirm that the setup of the Fast Clamp is on. (Refer to “Confirming the Setup” (page 48).)
51
• Do not attach the Fast Clamp to the blood inlet line of the centrifugal pump. Air bubbles may occur if the blood inlet line is
clamped, as the inside of the centrifugal pump will be negatively pressurized.
Warning
Attaching the Fast Clamp
• Do not attach the Fast Clamp to any tube other than the dedicated tube (internal diameter: 9.5 mm (3/8 inches)). If a tube
other than the dedicated tube is used, this may result in decreased performance.
Caution