Specifications
Dimensions,
unit weight
Temperature
Sensor Cable
Cable length 2.8 m
Pressure Sensor
Cable
Cable length 2.3 m
CDI
Communication
Cable
Cable length 2 m
AC power cable Cable length 3 m
Cart Dimensions 466 mm (width) × 1728 mm (maximum height) x 642 mm (depth)
* Excluding width and depth of caster
Unit weight 35 kg
Related components • CAPIOX centrifugal pump
• CAPIOX EBS Circuit with X coating
• CAPIOX Luer Thermistor
• CDI System 500
• Pressure sensor
(Compliant with ANSI/AAMI BP22:1994)
* Limited to items that can be connected to connectors.
Accessories
• AC power cable (1)
• Instructions for use (1)
• Backup kit (1)
<Contents>
• Backup Controller (1)
• Drive Motor (without pole clamp) (1)
• AC power cable (1)
• Screwdriver (+) (1)
• Extra screws (to fix the drive motor) (2)
• Instructions for use (Backup Controller) (1)
* See the instructions for use (Backup Controller) included in the backup kit for system specifications (Backup Controller).
The specifications and external appearance of the product may be changed without notice for the purpose of improvement.
Units Available
Catalogue number Plug type Voltage Socket type
ME
*
SP200C
American type 100 to 120 V
ME
*
SP200C2
UK type 100 to 240 V
ME
*
SP200C3
European type 100 to 240 V
140

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Terumo CAPIOX EBS Circuit with X coating and is the answer not in the manual?

Terumo CAPIOX EBS Circuit with X coating Specifications

General IconGeneral
BrandTerumo
ModelCAPIOX EBS Circuit with X coating
CategoryMedical Equipment
LanguageEnglish

Related product manuals