Operation Procedure
6. Close the lid of the Fast Clamp, and close the latch.
(Figure 1-61)
• The clamp ready lamp is lit in green.
Figure 1-61
Perform the inspection before circulation starts by following “Inspection Before Circulation Starts (Before Each Use)”. (Refer to
page 125.)
Starting Circulation
1. Confirm that the Motor speed indicator in the LCD or the
motor speed indicator in the control panel indicates “0”.
2. Clamp the blood outlet line of the centrifugal pump.
(Figure 1-62) Turn the motor speed knob to the coast
speed. Gradually release the clamp of the blood outlet line.
• For details about how to operate the centrifugal pump, refer to the instructions for
use of the centrifugal pump to be used.
3. While monitoring the flow, adjust the motor speed to
achieve the necessary flow.
Figure 1-62
55
Note
Check
Inspection Before Circulation Starts
• Do not use this system as a suction device. The system equipment is not intended to perform sucking operations.
• Use this system while paying attention not to let air in the blood circulation circuit. Health damages may occur to the
patient.
• When this system is used for negative-pressure blood removal aid, be cautious as the correlation between the flow and
motor speed of the centrifugal pump will change according to the strength of negative pressure applied to the reservoir.
• Do not let anesthetic drugs come into contact with the system or accessories. The system or accessories may be
damaged.
Cautions
Latch
Clamp ready lamp
Clamp
Blood outlet line