Operation Procedure
Prepare the following devices.
● Controller
● AC power cable
● Drive Motor
● ABD/Flow Sensor
● Applicable centrifugal pump
● Vaseline
● Applicable blood circuit (including oxygenator, etc.)
● Priming solution
Prepare the following devices if necessary.
● Cart
● Fast Clamp Arm
● Fast Clamp Attachment (for EBS)
● Fast Clamp
● Applicable temperature sensor (CAPIOX Luer Thermistor)
● Temperature Sensor Cable
● Applicable pressure sensor
● Pressure Sensor Cable
● Dedicated holder for applicable blood circuits
For instructions on attaching other products, refer to “Attaching Other Products to the Cart” (page 61).
1. Lock all casters of the Cart with each stopper. (Figure 1-2)
Figure 1-2
29
• Do not place the Controller on a soft surface such as a bed. The system may drop or fall. The air inlet (with filter) at the
bottom surface of the Controller may be closed.
• When the Cart is not being used, place it in a stable place.
Cautions
Preparing the Necessary Devices
Attaching the Controller to the Cart
Stopper
Lock