Various menu settings
3. Tap [OK] to complete the setting. (Figure 3-71)
Figure 3-71
• If you do not tap [OK], the changes to the setting will not take effect.
The admin setup of response behaviours (Message Only/Pump Coast/Clamp+Pump stop) for the Pressure Alarm, the Air Bubble
Detected Alarm and the Back Flow Alarm can be set up.
For details about how to set, refer to “Setting the Safety Connection” (page 91).
• The changes to the setting will take effect after restarting.
The on/off status of various functions, alarm or alert threshold values and the settings of the safety connections can be confirmed
in the list. Also, all admin setup values can be reset to the factory settings as follows.
For details about how to confirm the settings using the list, refer to “Alarm Setup List” (page 92).
This section describes the procedures for returning all the admin setup values to the factory settings.
Resetting to the factory settings
1. Tap [Setup for Shipment] on the Alarm Setup List screen.
(Figure 3-72)
Figure 3-72
2. Tap [OK]. (Figure 3-73)
The settings displayed in the Alarm Setup List screen will be reset to the
factory settings.
Figure 3-73
105
Note
Setting the Safety Connection
Note
Alarm Setup List

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Terumo CAPIOX EBS Circuit with X coating and is the answer not in the manual?

Terumo CAPIOX EBS Circuit with X coating Specifications

General IconGeneral
BrandTerumo
ModelCAPIOX EBS Circuit with X coating
CategoryMedical Equipment
LanguageEnglish

Related product manuals