Various menu settings
12. Tap [OK] to complete the setting. (Figure 3-18)
Figure 3-18
• If you do not tap [OK], the changes to the setting will not take effect.
• The lower/upper limits of the alarm cannot be input unless the values are higher/lower than those of the alert. When the
values cannot be input, check that the values of the lower/upper limits of the alarm are correct in relation to the lower/upper
limits of the alert.
The following settings are available:
• The on/off status of the temperature measuring function
• The threshold value of the Temperature Alert (setup range: 0.0 to 50.0°C)
1. Tap the Temperature Alarm icon. (Figure 3-19)
The Temperature Alarm screen opens.
2. Tap the channel ([T1] or [T2]) you want to set. (Figure 3-20)
The following figures show the example when [T1] is selected.
Figure 3-19
<<Setting the on/off status of the temperature measuring
function>>
3. Tap [On] or [Off]. (Figure 3-21)
When turning it off, tap [OK] to complete the setting.
Figure 3-20
Figure 3-21
88
Notes
Setting Temperature Measuring Function

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Terumo CAPIOX EBS Circuit with X coating and is the answer not in the manual?

Terumo CAPIOX EBS Circuit with X coating Specifications

General IconGeneral
BrandTerumo
ModelCAPIOX EBS Circuit with X coating
CategoryMedical Equipment
LanguageEnglish

Related product manuals