Various menu settings
When you input the patient
ʼs height and weight, you can check the calculated BSA (body surface area) value.
Subsequently, the perfusion index is displayed in the perfusion index indicator on the LCD touch panel display. For details about
the perfusion index indicator, refer to “LCD Touch Panel Display (When Operating Normally)” (page 9).
1. Tap the BSA icon. (Figure 3-38)
The BSA screen opens.
2. Tap or to input height and weight values.
(Figure 3-39)
The height and weight can be set within the ranges below.
Height: 40.0 to 250.0 cm
Weight: 1.0 to 300.0 kg
Figure 3-38
3. Tap [Dubois] or [Boyd] for the calculation method of BSA.
(Figure 3-40)
The BSA value is displayed in accordance with the selected calculation
method.
Figure 3-39
4. Tap [OK] to complete the setting. (Figure 3-41)
Figure 3-40
Figure 3-41
• If you do not tap [OK], the changes to the setting will not take effect.
93
Setting BSA
Note