Loading...
O
2
transfer
CO
2
transfer
Heat exchanger performance factor
Pump performance
SP45 SL45
Data
Blood Temperature:37±1
:Bovine Blood
:12±1g/dL
:7.4
:65±5%
:45±5mmHg
:0±5mEq/L
:1.0
<Conditions>
Blood
Hb
pH
SvO2
PvCO2
B.E.
V/Q
<Conditions>
Blood
Hb
pH
SvO2
PvCO2
B.E.
V/Q
:Bovine Blood
:12±1g/dL
:7.4
:65±5%
:45±5mmHg
:0±5mEq/L
:1.0
Blood Temperature:37±1
<Conditions>
Blood
Hb
Tbi
Twi
Water flow rate
:Bovine Blood
:12±1g/dL
:30±1
:40±1
:15 L/min
Blood flow rate(L/min)
8 6 4 2 0
500
400
300
200
100
0
500
400
300
200
100
0
0
2
4
6
8
Blood flow rate
(
L/min
)
O
2
trandfer rate(mL/min)
CO
2
trabsfer rate(mL/min)