Operation Procedure
3. While holding the lid of the Fast Clamp, pull up the lever
until you hear a click. (Figure 1-58)
The plunger is set at the open position.
To close the plunger manually, use [Clamp] button on the Fast clamp screen.
(Refer to page 57.)
• After pulling up the lever, the lever has returned to the original position.
• If the lever does not return to the original position, pull down the lever. If it does
not return to the original position, a failure may have occurred.
Stop using the system immediately, and contact Terumo’s service representative.
4. Release the latch of the Fast Clamp lid, and open the lid.
(Figure 1-59)
Figure 1-58
5. Insert the tube for the blood line into the Fast Clamp.
(Figure 1-60)
Figure 1-59
• When inserting the tube, avoid excessive bending which may degrade the flow.
Figure 1-60
54
Note
Check
Note
Lever
Latch
Tube