Maintenance and Inspections
4. Plug in the AC power cable.
• The On Battery Alarm stops.
• The AC power lamp is lit.
• The AC power indicator changes to .
5. Change the motor speed to 0 RPM to stop the Drive Motor.
6. Turn off the power.
Inspection method of the Drive Motor Disconnected Alarm function and the ABD/Flow
Sensor Disconnected Alarm function (Before each use)
1. Disconnect the plug of the Drive Motor and turn on the
power.
• “Drive Motor Disconnected” is displayed on the message area and an alarm
sounds. (Figure 5-2)
2. Tap . (Figure 5-3)
• The alarm is muted.
Figure 5-2
3. Turn off the power and connect the Drive Motor.
4. Disconnect the plug of the ABD/Flow Sensor and turn on
the power.
Figure 5-3
• “ABD/Flow Sensor Disconnected” is displayed on the message area and an alarm
sounds. (Figure 5-4)
5. Tap . (Figure 5-5)
• The alarm is muted.
Figure 5-4
6. Turn off the power and connect the ABD/Flow Sensor.
Figure 5-5
124
Check
Checks
Check
Check
Check

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Terumo CAPIOX EBS Circuit with X coating and is the answer not in the manual?

Terumo CAPIOX EBS Circuit with X coating Specifications

General IconGeneral
BrandTerumo
ModelCAPIOX EBS Circuit with X coating
CategoryMedical Equipment
LanguageEnglish

Related product manuals