Maintenance and Inspections
Inspection method of the Fast Clamp (Before each use)
When using the Fast Clamp, follow the procedures below to perform the inspection:
1. Attach the tube of the blood line to the Fast Clamp.
2. Tap the setup icon. (Figure 5-8)
3. Tap the Fast clamp icon on the Setup screen. (Figure 5-9)
The Fast clamp screen opens.
Figure 5-8
4. Tap [Clamp] to operate the Fast Clamp. (Figure 5-10)
Figure 5-9
The plunger is closed (up position).
5. Tap [Cancel] to close the screen. (Figure 5-11)
Figure 5-10
Figure 5-11
126
Check