Alarms and Troubleshooting
Occurrence Cause Action
The temperature is displayed as
“--.-”.
The temperature is not stable.
The Temperature Sensor Cable is not
properly connected.
Reconnect the Temperature Sensor Cable.
(Refer to page 37.)
The CAPIOX Luer Thermistor is not
properly connected.
Reconnect the CAPIOX Luer Thermistor.
(Refer to page 46.)
The CAPIOX Luer Thermistor has failed. Replace the CAPIOX Luer Thermistor.
The Temperature Sensor Cable has failed. Contact Terumo’s service representative.
The Clamp ready lamp does not
light up.
The Fast Clamp is not properly connected. Connect the Fast Clamp. (Refer to
page 38.)
The fast clamp function is not on. Enable the fast clamp function. (Refer to
page 90.)
The safety connection which allows the
Fast Clamp to operate has not been set.
Set the safety connection. (Refer to
page 91.)
The tube has not been securely set.
The lid is not properly closed.
Set the tube and close the lid. (Refer to
page 51.)
The Fast Clamp cannot be
clamped from the button on Fast
clamp screen.
The tube has not been securely set.
The lid is not properly closed.
Set the tube and close the lid. (Refer to
page 51.)
The fast clamp function is not on. Enable the fast clamp function. (Refer to
page 90.)
The Drive Motor is running. Execute the clamp after stopping the Drive
Motor.
The key lock function does not
activate.
An alarm other than the On Battery Alarm
is issued.
Resolve the cause of the alarm. (Refer to
page 134.)
Log data is being exported. The key lock function does not activate
while log data is being exported. This is
not a malfunction.
The count-down ended sound
does not sound.
An alarm is issued. When an alarm is issued, the alarm sound
has priority over the count-down ended
sound. This is not a malfunction.
The button that exports history
data to a USB memory drive does
not function.
The Drive Motor is running.
Output data to the USB memory drive after
stopping the Drive Motor.
“Failed to output the log to the
USB.” is displayed.
The USB memory drive cannot be written. Enable the USB memory drive writable,
and try to export the data again.
A USB memory drive is not
recognized.
Removal of the previous USB memory
drive was not properly executed.
Turn off the power of the Controller, and
turn on the power again.
Enable the USB memory drive writable,
and try to export the data again.
The password to open admin
setup is lost.
Contact Terumo’s service representative to
initialize the password.
133

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Terumo CAPIOX EBS Circuit with X coating and is the answer not in the manual?

Terumo CAPIOX EBS Circuit with X coating Specifications

General IconGeneral
BrandTerumo
ModelCAPIOX EBS Circuit with X coating
CategoryMedical Equipment
LanguageEnglish

Related product manuals