Precautions
Cautions
(Hand Crank)
• When using the Hand Crank, attach it to a pole securely after ensuring there is enough space and tube length so that the
crank can rotate smoothly. [It may become impossible to maintain circulation.]
• Do not exceed the maximum motor speed of 3000 RPM. [This may cause problems to the centrifugal pump or blood
damages may occur.]
• Do not place heavy items on the flexible shaft or do not bend the shaft by applying excessive force.
(Drive Motor)
• When attaching the centrifugal pump to the Drive Motor, be careful not to pinch your finger. [Your finger may be injured.]
• Do not hang cables (the Drive Motor, the ABD/Flow Sensor) or AC power cable on the fixed hook. [The fixed hook may be
damaged.]
• When the centrifugal pump is not used, attach the magnet cover to the pump receptacle of the Drive Motor.
(ABD/Flow Sensor)
• When using the ABD/Flow Sensor, use it after cleaning the detection window. [It may not correctly detect flow and air
bubbles.]
• When the air bubble detector function turns on, pay attention to the handling of the blood circuit and the ABD/Flow Sensor.
[If a physical shock occurs to the blood circuit or to the ABD/Flow Sensor, or if the latch is touched carelessly resulting in
the detachment of the ABD/Flow Sensor from the tube, the Air Bubble Detected Alarm will be issued.]
• When the air bubble detection sensitivity is high, attach the sensor so that the ABD/Flow Sensor detection window and the
direction of gravitational force are orthogonal. [It may not correctly detect air bubbles.]
(Cart)
• Keep the cartʼs casters in a locked position except when transporting the cart. [The Cart may move unintentionally,
causing attached products to fall.]
• Use both hands to grasp the cart
ʼs handles. [If it is not held properly, the cart may move in an unexpected direction and
risk patient injury.]
• Do not put any objects other than those specified for use with the system at the upper part of the Cart. Do not lean on the
Cart. [The Cart may fall or topple.]
• When the Cart is not being used, place it in a stable place. [This system may fall leading to failure.]
24

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Terumo CAPIOX EBS Circuit with X coating and is the answer not in the manual?

Terumo CAPIOX EBS Circuit with X coating Specifications

General IconGeneral
BrandTerumo
ModelCAPIOX EBS Circuit with X coating
CategoryMedical Equipment
LanguageEnglish

Related product manuals