Calibration and verification schedule
The instrument is factory-calibrated and does not require calibration at installation. To
ensure optimal performance, perform calibrations at the recommended frequency.
Note: After instrument installation, we recommend performing instrument
verication using the provided RNase P plate.
IMPORTANT! Perform calibrations and instrument runs under the environmental
conditions that are specied in “Environmental requirements“ on page 89. Exposure
to extreme temperatures can adversely aect the instrument performance and shorten
the life span of the instrument components.
To set the calibration frequency for the instrument, touch Seings4Maintenance
and Service4Calibrations4History and Reminders4Edit4Exp interval eld.
Calibration
Recommended frequency
ROI/Uniformity
• Every two years (recommended)
• Always perform new Background and Dye calibrations after
an ROI/Uniformity calibration.
Note: Performing an ROI/Uniformity calibration invalidates all
other calibrations.
5
QuantStudio
™
1 Real-Time PCR System Installation, Use, and Maintenance Guide