During the initial startup, you will be automatically prompted to select a Cloud
region. Return to the selection through Seings, if necessary.
1.
In the home screen, touch Seings4Instrument Seings4Cloud Region.
2.
Touch a region.
The server used to store your data is located in this region.
3.
Touch OK.
1.
In the home screen, touch
Seings4Instrument Seings4Date/Time.
2.
Select a time zone from the dropdown list.
3.
Select a date format.
a.
Touch Date Format, then select the preferred date format.
b.
Touch Next, touch the date eld, then enter the date.
c.
Touch Enter, then touch Done.
4.
Select a time format.
a.
Touch Time Format.
b.
Slide the control to select a 12-hour or 24-hour clock.
c.
Touch Next, touch the time eld, then enter the time.
d.
Touch Enter, then touch Done.
5.
Touch Done.
IMPORTANT! Congure the instrument for either wired connection or wireless
connection but not both.
“Set up a wired connection“ on page 33
“Set up a wireless connection“ on page 33
Select a Cloud
region
(Administrator
only)
Set the date and
time
(Administrator
only)
Configure the
network
Chapter 3 Configure the instrument and manage instrument profiles
Manage instrument settings
3
32
QuantStudio
1 Real-Time PCR System Installation, Use, and Maintenance Guide